AC HI-POT SPECIFICATIONS
|
Voltage Range |
0.100 ~ 5.000kV |
Voltage Resolution |
5V |
Voltage Regulation |
1% + 5 V |
Voltage Accuracy |
±1% of reading ±10 V (above 500V) |
Max. Current |
40mA, (above 500V) |
Current Limit |
0.10 ~ 40mA, 0.02mA (resolution) |
Current Accuracy |
±1% of reading ±80μA |
DC HI-POT SPECIFICATIONS
|
Voltage Range |
0.100 ~ 6.000kV |
Voltage Resolution |
5V |
Voltage Regulation |
1% + 5 V |
Voltage Accuracy |
±1% of reading ±10 V (above 500V) |
Max. Current |
10mA, (above 500V) |
Current Limit |
0.10~10mA, 0.01mA (resolution) |
Current Accuracy |
±1% of reading ±50μA |
INSULATION RESISTANCE SPECIFICATIONS
|
DC Voltage |
50V/100V/500V/1000V |
Resistance Range |
1 ~ 9900M(500V,1000V); 1 ~ 2000M(50V,100V) |
Resistance Accuracy |
50V/100V 1~50MΩ: ±5% of reading ±1Count 51~200MΩ: ±10% of reading ±1Count 201M~1990MΩ: ±20% of reading ±1Count |
|
500V/1000V 1~500MΩ: ±5% of reading ±1Count 501~2000MΩ: ±10% of reading ±1Count 2001M~9900MΩ: ±20% of reading ±1Count |
GROUND BOND SPECIFICATIONS
|
Voltage Range |
Max. 8V (DC) |
Current Range |
3 ~ 42A |
Current Resolution |
0.02A |
Current Accuracy |
1% of reading 0.2A(3A~8A) |
Ground Bond |
1% of reading 50mA(9A~42A) |
Resistance Test |
0 ~ 620mΩ |
Resistance Resolution |
0 ~ 600mΩ(100μΩ) |
Resistance Accuracy |
±1% of reading 3 mΩ |
Test Method |
Four terminals test mode |
ARC DETECT
|
Detect Current |
40 level (1 ~ 40mA) |
INTERFACE
|
Remote I/O |
Standard |
RS-232C |
Standard |
Scan |
Standard |
Real Panel Output |
Standard |
GPIB |
Option |
Scanner Box |
Option (HSB-001-1 : 8ch HV+8ch GB ; HSB-001-2 : 16ch HV) |
POWER SOURCE |
AC 100V, 120V, 220V, 230V ± 10%, 50/60Hz |
|
|
ACCESSORIES
|
|
Instruction manual x 1 Power cord x 1 Test leads GHT-105A x 1, GTL-115 x 1 |
DIMENSIONS & WEIGHT |
330(W) x 149(H) x 446(D) mm Approx.15kg |